Abstracts
of the December 2007 Issue
Working time of elastomeric
impression materials:
Relevance of rheological tests
Markus Balkenhol,
dds
, Masafumi Kanehira,
dds
, phd, Werner
J. Finger,
dds
, phd
& Bernd
Wöstmann,
dds
Abstract: Purpose:
To study possible relationships between rheological parameters
of elastomeric impression materials and their dimensional
accuracy at different stages during setting within the
manufacturers’ recommended working time. Methods:
Four chemically different impression material-types were
tested. Impressions were taken of a master cone at 30-second
intervals after mixing and up to 30 seconds beyond the
recommended working time. Impression accuracy (Dd)
was determined as the dimensional deviation of stone dies from
the master cone. The phase angle d(t)
and storage modulus G’(t) were measured and the working time
determined according to ISO 4823. Three operators assessed the
latest point in time for uncompromised impression taking.
Correlation analysis was performed using Spearman’s
Rho
(P= 0.05). Results:
For most materials investigated, Dd(t) was constant within the manufacturers’ recommended working time
whereas d(t)
and G’(t) changed significantly. There was no correlation
(P> 0.05) between Dd(t)
versus d(t)
and G’(t), respectively. When determined according to ISO
4823, working time was longer than operator-assessed working
time for all materials investigated. (Am
J Dent 2007;20:347-351).
Clinical significance:
Dentists should preferably take impressions reasonably early
within the manufacturers’ recommended working time in
order to avoid inaccuracy when approaching the very end of
the working time. Rheological tests, including the procedure
described in the ISO 4823, are unsuitable for identifying
the latest point in time at which impression taking is still
possible with consistent accuracy using the impression
materials tested.
*:
Dr. Markus Balkenhol, Schlangenzahl 14, 35392 Giessen,
Germany.
E-*:
markus.balkenhol@dentist.med.uni-giessen.de
Post-endodontic
restorations with adhesively luted fiber-reinforced
composite
post systems: A review
Kerstin
Bitter, dr
med
dent
&
Andrej M.
Kielbassa, dr
med
dent
Abstract: Purpose: To review the literature on adhesive luting of
fiber-reinforced composite posts (
FRC
) to provide evidence for the clinical procedure of restoring
endodontically treated teeth using
FRC
posts. Methods:
Data focusing on bonding behavior between root canal dentin,
luting agent, and
FRC
post in vitro as
well as in vivo
performance of teeth restored with
FRC
posts were reported. These data were identified by searches of
“PubMed”, “Scopus”, and “Cochrane Library”
databases with the terms “post-endodontic restoration”,
“fiber post”, “adhesive luting”, “root canal
dentin”, “clinical study”, and “pre-treatment fiber
post”. Papers published up to September 2007 were selected,
and most relevant references were chosen. Cross-referencing of
significant papers identified additional relevant articles. Results:
FRC
posts seem to have become increasingly popular for the
restoration of endodontically treated teeth. Compared to metal
posts,
FRC
posts revealed reduced fracture resistance in
vitro, along with a usually restorable failure mode.
Bonding behavior among
FRC
post, luting agents, and root canal dentin demonstrated
varying results. Bond strengths between
FRC
posts and resin cements can be enhanced by using various
pre-treatment procedures; however, bonding to root canal
dentin still seems to be challenging. Most clinical studies
investigating survival rates of teeth restored with
FRC
posts revealed promising results, but risk factors (e.g.,
the loss of coronal tooth structure) have not been studied
intensively. In addition, randomized controlled clinical long
term trials are scarce. (Am
J Dent 2007;20:353-360).
Clinical significance:
Evidence-based recommendations are still not possible, and
further research using standardized protocols is warranted.
*:
Dr. Kerstin Bitter
, Department of Operative Dentistry and Periodontology, Charité
Centrum 3 for Dental Medicine, Charité - Universitätsmedizin
Berlin, Aßmannshauser Straße 4-6, D - 14197 Berlin, Germany.
E-*:
kerstin.bitter@charite.de
Five-year
clinical evaluation of One-Up
Bond F in non-carious
cervical
lesions
Michael
F. Burrow,
mds
, phd,
med
&
Martin J.
Tyas,
bds
, phd ddsc
Abstract:
Purpose: To
evaluate the retention and marginal staining of restorations
placed in non-carious cervical lesions using the all-in-one
self-etching system, One-Up Bond F and Palfique Estelite resin
composite. Methods: Fifty-one non-undercut non-carious cervical lesions
received restorations with One-Up Bond F and Palfique Estelite
in 15 subjects (mean age 57.7 years) by one operator. Teeth
were cleaned with pumice and water, restored following the
manufacturer's instructions and evaluated at 6 months, 1, 2, 4
and 5 years for retention and presence of marginal staining. Results:
At 1 year, 42 restorations were evaluated, and all were intact
with three restorations showing slight marginal staining. At 2
years, 37 restorations were evaluated from 11 subjects, and
all were intact. Marginal staining was observed around nine
restorations. At 4 years, 27 restorations and at 5 years, 24
restorations could be evaluated. Marginal staining showed a
significant increase at the enamel and cervical margins over
the 5 years of the study, with two restorations showing
severe, nine mild and 13 no staining. Although a number of
restorations could not be evaluated, the retention rate of 92%
at 5 years showed One-Up Bond F/Palfique Estelite resin
composite for the restoration of non-carious cervical lesions
produced an excellent outcome. Marginal staining was limited
to a small number of subjects and is unlikely to be a problem
if margins are refurbished at recall visits. (Am
J Dent 2007;20:361-364)
Clinical significance: The
all-in-one adhesive, One-Up Bond F, with Palfique Estelite
resin composite is an effective adhesive for restorations in
non-carious cervical lesions.
*:
Dr. Michael F. Burrow,
School
of
Dental
Science,
720 Swanston Street
,
Melbourne
3010,
Victoria
,
Australia
. E-*:
mfburrow@unimelb.edu.au
Susceptibility
of a polycaprolactone-based root canal filling material to
degradation. Evidence of biodegradation from a simulated field
test
Franklin R. Tay, bdsc (hons), phd, David
H. Pashley, dmd, phd,
Robert
J. Loushine,
dds
, Sergio Kuttler,
dds
, Franklin García-Godoy,
dds
, ms, Nigel
M. King,
bds
, ms, phd
& Marco
Ferrari, md,
dds
, phd
Abstract:
Purpose: To examine
if Resilon, a polycaprolactone-based root filling material,
was susceptible to microbial biodegradation by using a
simulated field test that consisted of incubating the material
in wet dental sludge under mesophilic and aerobic conditions. Methods:
Pressed disks prepared from Resilon, polycaprolactone
(positive control) and gutta-percha (negative control) were
incubated in wet dental sludge for up to 4 months and examined
for topographical changes using scanning electron microscopy. Results:
Gutta-percha exhibited minimal changes in surface integrity,
while polycaprolactone and Resilon exhibited severe surface
pitting and erosion. In the latter, disappearance of the
polymer matrix was accompanied by exposure of mineral and
bioactive glass fillers. Bacteria and hyphae-like structures
were present on the disk surfaces. (Am
J Dent 2007;20:365-369).
Clinical significance: The incorporation of
polycaprolactone in the Resilon polymer/filler blend did not
prevent its microbial biodegradation in a simulated field
study. Thus, the potential for clinical biodegradation of
Resilon in root canals should be further investigated in
long term clinical trials using methodologies other than
radiographic examination alone, as fillers do not undergo
biodegradation and will appear radiopaque on radiographs
even after the degradation of a radiolucent polymeric
component.
*: Dr. Franklin Tay, Department of Oral Biology and Maxillofacial
Pathology, School of Dentistry,
Medical
College
of Georgia,
Augusta
,
GA
,
30912-1129
,
USA
. E-*:
tayfranklin7@gmail.com
The effect of direct and indirect water storage on the microtensile dentin
bond strength of a total-etch and two self-etching adhesives
Ali
I. Abdalla,
dds
, phd, Magda
El Eraki,
dds
, phd
& Albert
J. Feilzer,
dds
, phd
Abstract:
Purpose: To
evaluate the effect of direct and indirect water storage on
the microtensile dentin bond strength of one total-etch and
two self-etching adhesives. Methods:
The adhesive materials were: one total-etch adhesive; Admira
Bond and two self-etch adhesives; Clearfil SE Bond and Hybrid
Bond. Freshly extracted human third molar teeth were used. In
each tooth, a Class I cavity (4 x 4 mm) was prepared in the
occlusal surface with the pulpal floor extending approximately
1 mm into dentin. The teeth were divided into three groups
(n=12). Each group was restored with the resin composite
Clearfil APX using one of the tested adhesives. For each
experimental group three test procedures (n=10) were carried
out: Procedure A: the teeth were stored in water for 24 hours,
then sectioned longitudinally, buccolingually and
mesiodistally to get rectangular slabs of 1.0 – 1.2 mm
thickness on which a microtensile test was carried out;
Procedure B: the teeth were also sectioned, however the slabs
were stored in water at 37°C for 1 year before microtensile
testing; Procedure C: the teeth were kept in water at 37°C
for 1 year before sectioning and microtensile testing. During
microtensile testing the slabs were placed in a universal
testing machine and load was applied at a cross-head speed of
0.5 mm/minute. Results: For the 24-hour water storage groups there was no
significant difference in bond strength between the three
adhesives. After 1 year of indirect water storage, the bond
strength decreased but the reduction was not significantly
different from those of 24 hours. After 1 year of direct water
storage, the mean bond strengths of Admira Bond and Hybrid
Bond were significantly reduced compared to their 24-hour
results. In contrast the average value of Clearfil SE Bond was
not significantly affected. (Am
J Dent 2007;20:370-374).
Clinical
significance: All
the tested adhesives showed no reduction in bond strength
after indirect water exposure for 1 year. After 1 year
direct water exposure, the bond produced by two-step
self-etch adhesive was stable. In contrast, the two-step
total-etch and all-in-one self-etch adhesive were unable to
resist deterioration.
*: Dr. Ali I. Abdalla, Department of Restorative Dentistry, Faculty of
Dentistry,
University
of
Tanta
,
Tanta
,
Egypt
. E-*:
Aliebraheem01@yahoo.com
Surface treatments on quartz fiber post: Influence
on adhesion
and flexural properties
Nicolas Cheleux,
dds
, Patrick
Sharrock, phd
& Michel
Degrange,
dds
, phd
Abstract: Purpose: To evaluate (1) the effect of various surface treatments
to optimize post adhesion, and (2) if these surface treatments
had any adverse effect on the overall mechanical and chemical
properties of the posts. Methods: Six experimental
batches were set according to the nature of the surface
treatments which were followed or not by post silani-zation.
Post surface treatments were: control, chloroform (CHCl3)
and sandblasting (sand) with Al2O3 50 µm.
The silane (sil) used was Silane Coupling Agent Calibra. In
all cases, a dual-cure bonding agent was applied
Prime&Bond NT + Self Cure Activator. The Groups were: 1
(control), 2 (sil), 3 (CHCl3), 4 (CHCl3
+ sil), 5 (sand), 6 (sand + sil). Following surface
treatments, the posts were embedded in resin composite and the
composite-post interfacial strength was investigated with a
push-out test. Flexural and fatigue tests were performed to
control the mechanical performances of the treated posts.
Push-out data were analyzed using two-way ANOVA and Scheffé’s
multiple means comparisons test (α= 0.05).
SEM
observations revealed the topographical modifications induced
by sandblasting and the solvent. Results: Sandblasting
or surface treatment with chloroform increased the bond
strength (P< 0.001). The best values were obtained with the
combination of sandblasting followed by post silanization. The
flexural properties were not modified by the surface
treatments (P= 0.072) and all the posts reached 3,000,000
cycles without breaking. (Am J Dent 2007:20:375-379).
Clinical significance:
A combination of sandblasting followed by silane and bonding
agent application can be used in daily practice for
increasing post retention to the composite core without
modifying the mechanical resistance of the quartz fiber
post.
*: Dr. Nicolas Cheleux,
Faculty of Odontology, Paul Sabatier University, 3 chemin des
Maraîchers, 31062 Toulouse,
France
. E-*:
cheleux@club-internet.fr
Marginal
accuracy of press-ceramic veneers influenced by preparation
design and fatigue
Christian
F.J. Stappert,
dds
, ms, dr
med
dent
, Ummuhan
Ozden,
dds
, dr
med
dent
, Wael Att,
dds
, dr
med
dent
, Thomas Gerds, dipl-
math
, dr rerr
nat
&
Joerg R.
Strub,
dds
, dr
med
dent
, phd
Abstract:
Purpose: To
investigate the influence of preparation design and mouth
motion fatigue on the marginal accuracy of press-ceramic
veneers bonded to human maxillary central incisors. Methods:
Forty-eight extracted human maxillary central incisors divided
in three groups of 16 specimens each, received veneer
preparations. In Group WP a window preparation (WP) was
performed. Specimens in Group IOP were prepared with an
incisal overlap (IOP) of 2 mm without palatal chamfer. For
Group CVP, specimens received a complete-veneer preparation
(CVP) of 3 mm incisal reduction and 2 mm palatal extension.
Veneers were fabricated with
IPS
Empress and bonded adhesively with dual polymerizing composite
Variolink II. The discrepancies of marginal fit were examined
on epoxy replicas before and after mouth motion fatigue at
x200 magnification. Results: The mean (geometrical) (95% confidence limits) marginal
accuracy after cementation was recorded as follows: WP-46
(43-49) µm, IOP-46 (44-49) µm and CVP-54 (51-58) µm.
Fatigue led to a significant decrease of marginal accuracy in
all groups (P< 0.036) [WP-47 (44-50) µm, IOP-51 (48-53) µm
and CVP-63 (59-67) µm]. Group CVP demonstrated significantly
higher marginal gap values than groups WP and IOP, before and
after fatigue (P< 0.004). No significant differences were
found between groups WP and IOP. (Am
J Dent 2007;20:380-384).
Clinical significance: Examined preparation designs and
luting method demonstrated a high marginal accuracy of
press-ceramic veneers in
vitro. The observed increase of marginal discrepancy for
complete press-ceramic veneers was considered to be
acceptable also for clinical application.
*:
Dr. Christian Stappert, Department of Biomaterials and
Biomimetics, New York University College of Dentistry, Arnold
and Marie Schwartz Hall of Dental Sciences, 345 East 24th
Street (Rm. 846S),
New York
, 10010 NY,
USA
. E-*:
christian.stappert@nyu.edu
Remineralization
of artificial interproximal carious lesions
using a fluoride mouthrinse
Markus
Jörg Altenburger,
dds
, Jörg
Fabian Schirrmeister,
dds
, Karl-Thomas
Wrbas,
dds & Elmar Hellwig,
dds
, phd
Abstract:
Purpose:
To evaluate remineralization and fluoride uptake of
demineralized enamel specimens in artificial interproximal
spaces using a 250 ppm fluoride mouthrinse (100 ppm amine
fluoride and 150 ppm sodium fluoride) compared to a
placebo-mouthrinse. Methods: This was a two-way, double-blind, in situ study. 24 volunteers wore intraoral appliances in the lower
jaw with mounted demineralized enamel specimens forming an
artificial interproximal space. The volunteers rinsed their
mouths for 28 days, twice a day with the respective mouthrinse
and brushed their teeth at least twice a day with a
fluoride-free dentifrice. The specimens were cleaned once a
day to simulate the use of dental floss. For analysis methods
transversal microradiography, fluoride content measurement and
quantitative light-induced fluorescence were used. Results:
Mineral gain was significantly higher after the use of the
fluoride mouthrinse: 571.88 Vol.%∙µm (±79.33
Vol.%∙µm
SEM
) compared to the placebo treatment 51.14 Vol.%∙µm (±72.14
Vol.%∙µm
SEM
). Significantly more fluoride uptake was observed after using
the fluoride mouthrinse: 3666.39 µg/cm³ (±513.50 µg/cm³
SEM
) than after the placebo mouthrinse: 516.95 µg/cm³ (±75.83
µg/cm³
SEM
). The quantitative light-induced fluorescence showed a
significant remineralizing effect after the fluoride
treatment: -388.47 %∙mm² (± 18.76%∙mm²
SEM
) compared to the placebo treatment -27.88%∙mm² (±20.19%∙mm²
SEM
). (Am J Dent
2007;20:385-389).
Clinical
significance:
The results demonstrated a clear remineralizing effect of
the fluoride mouthrinse tested and might be recommended for
caries prevention and during therapeutic phases to support
remineralization especially in difficult-to-reach areas.
*:
Dr. Markus Jörg
Altenburger, Department of Operative Dentistry and
Periodontology, Dental School and Hospital, Albert-Ludwigs
University Freiburg, Hugstetter Str. 49, 79095 Freiburg,
Germany. E-*:
markus.altenburger@uniklinik-freiburg.de
Influence of three different sealants on root
dentin demineralization in situ
Christian Ralf
Gernhardt, dr
med
dent
, Katrin
Bekes, dr
med
dent
,
dds &
Hans
-Guenter Schaller,
dr
med
dent
,
dds
Abstract: Purpose: To evaluate in situ the onset of initial
demineralization in human dentin pretreated either with two
different dentin bonding agents or a desensitizer. Methods:
28 freshly extracted human molars were included in this study.
The root surfaces were thoroughly cleaned, thereby removing
the cementum. From each tooth four root dentin specimens were
prepared. The specimens were distributed among the following
experimental groups: C: control group (untreated), S: Syntac
Classic, X: Xeno
III
, and H: Hyposen. For each subject, one dentin specimen of
each group was inserted into both buccal aspects of a
bilateral intraoral mandibular appliance. The appliances were
worn by 14 subjects for 5 weeks day and night. One side was
brushed daily with fluoride-containing toothpaste (Aronal)
(B). On the other side, plaque was allowed to grow (NB).
Individual oral hygiene techniques were performed without any
fluorides. During meals, the appliance was stored in 10%
sucrose solution. After the in situ period, slabs (150
µm) were ground and studied using polarized light microscopy.
Results: For lesion depth, ANOVA revealed significant
differences between brushed and unbrushed specimens. In the
brushed groups, the following
lesion depths were evaluated (mean values and standard
deviation in microns): Group B-C: 74.1 (±18.5), Group B-S:
27.6 (±9.6); Group B-X: 28.7 (±9.6), and Group B-H: 34.3 (±20.6).
The non-brushed specimens showed following lesion depths:
Group NB-C: 101.7 (±23.9), Group NB-S: 59.9 (± 13.1), Group
NB-X: 52.5 (± 12.1), and Group NB-H: 72.9 (± 19.9). Compared
to the non-brushed groups, lesion depths in the brushed group
were significantly decreased (P< 0.05, Tukey’s test). The
reduction of lesion depths after application of the three
agents was significant in both cases (brushed and non-brushed
groups). Within the limitations of an in situ study it
can be concluded that the demineralization of the root surface
can be inhibited by application of dentin adhesives and
desensitizers under different oral hygiene conditions. (Am
J Dent 2007;20:390-393).
Clinical significance:
The application of clinically proven dentin bonding agents
and desensitizing agents might have a caries-protective
effect on exposed root surfaces.
*: Dr. Christian Ralf Gernhardt, Department of Operative
Dentistry and Periodontology, University School of Dental
Medicine, Martin-Luther-University Halle-Wittenberg, Grosse
Steinstrasse 19, 06108 Halle, Germany. E-*:
christian.gernhardt@medizin.uni-halle.de
In
vitro wear of primary and permanent enamel.
Simultaneous erosion and abrasion
Gisele Maria
Correr,
dds
, ms, Roberta
Caroline Bruschi
Alonso,
dds
, ms, Simonides
Consani,
dds
, ms, phd,
Regina Maria
Puppin-Rontani,
dds
, ms, phd
& Jack
L. Ferracane,
phd
Abstract:
Purpose: To
evaluate the wear differences between primary and permanent
enamel caused by the association of abrasion and erosion. Methods:
Fragments of 40 human teeth (20 primary and 20 permanent) were
cast in acrylic rings, polished to expose a flat enamel area,
and evaluated with a 3D profiler. Antagonists were made from
primary and permanent molars. The specimens were distributed
into four groups (n=10) according to type of substrate and
slurry (neutral and acidic), and cycled 100,000 times in the
OHSU oral wear simulator. Specimens were cleaned and
re-profiled. Volume loss and maximum depth were determined on
the polished specimens. Data were submitted to ANOVA and
Tukey’s test. The area of the wear facet on the antagonist
was measured. Results:
Primary enamel showed more wear than permanent enamel,
regardless of the slurry type. The acidic environment
increased the wear rate compared to the neutral slurry. The
association of abrasion and erosion increased the wear rate
for the primary teeth only. There was no significant
difference among groups for cusp wear and no correlation
between cusp wear and the enamel wear. (Am J Dent 2007;20:394-399).
Clinical significance:
Primary teeth are more susceptible to wear than permanent
teeth, especially in an acidic environment, and limited
exposure to acidic food and drink should be recommended to
prevent accelerated tooth wear in children.
*: Dr. Gisele Maria Correr, Department of Restorative Dentistry, Dental
Material Area, Piracicaba Dental School, UNICAMP, Av. Limeira,
901, CEP 13414-018, Piracicaba, SP, Brazil.
E-*:
giselefop@yahoo.com
Cytotoxic effects
of different concentrations of chlorhexidine
Lorena Brito
de Souza,
dds
, Sabrina
García de Aquino,
dds
, Pedro
Paulo Chaves
de Souza,
dds
, ms, phd, Josimeri Hebling,
dds
,
ms
, phd & Carlos
Alberto de Souza Costa,
dds
,
ms
, phd
Abstract:
Purpose: To
evaluate the cytotoxic effects of different concentrations of
Chlorhexidine (Chx) to the odontoblast cell line MDPC-23. Methods:
The odontoblast-like cells were seeded (30,000 cells/cm2)
in 60 wells of 24-well dishes and then incubated in contact
with the following experimental and control solutions: Group
1: 0.0024% Chx; Group 2: 0.004% Chx; Group 3: 0.02% Chx; Group
4: Phosphate buffer saline solution (PBS, negative control);
and Group 5: 0.06% H2O2 (positive
control). Cell metabolic activity was measured by MTT assay
and the cell morphology was analyzed by
SEM
. Results: The
cytotoxic effects of Chx are dose-dependent. The reduction in
the cell metabolism for Groups 1, 2, and 3 was 24.8%, 29.9%
and 70.8%, respectively. No statistical difference was
observed between the Groups 1 and 2 in which no significant
cell morphology changes occurred. Consequently, it was
concluded that 0.02% Chx solution presents high cytotoxicity
to the odontoblast-like cells MDPC-23. On the other hand,
0.0024% and 0.004% Chx causes slight cytopathic effects to the
cultured cells. (Am J
Dent 2007;20:400-404).
Clinical
significance:
Chlorhexidine (Chx) solutions do not interfere with
the bonding mechanism during adhesive restoration and
inhibit further degradation of the resin/dentin interface.
In addition, low concentrations of Chx, such as 0.0024% and
0.004% do not cause significant toxic effects to the pulp
cells and can be regarded as a potential chemical agent to
be used as a cavity cleanser.
*:
Dr. Carlos Alberto de Souza Costa, University of São Paulo
State / UNESP, Rua Humaitá 1680, CEP: 14.801-903, CP: 331 –
Centro, Araraquara, SP,
Brazil
. E-*:
casouzac@foar.unesp.br
Fracture load of fixed partial dentures anchored by
composite inlays
Brigitte
M. Ohlmann, dr
med
dent
, Marc
Schmitter, pd, dr
med
dent
, Olaf
Gabbert, dr
med
dent & Peter
Rammelsberg, prof, dr
med
dent
Abstract:
Purpose:
To evaluate in vitro
the fracture load of fixed partial dentures (FPDs) anchored by
use of composite inlays. The effects of span length,
silica-coating, mechanical loading and framework material were
also tested. Methods: Defined
box inlay cavities were made on a mandibular molar and a
premolar. Fifty-six FPDs were manufactured using a polymer
composite material and received an industrial prefabricated
polymethylmethacrylate (PMMA) frame for stabilization. All
FPDs underwent thermal cycling. The FPDs were divided into
test groups simulating the effect of different span length (7
or 12 mm), tribochemically silicoating (yes or no) and
mechanical loading (yes or no). The load to fracture was
measured and fracture sites were evaluated. As a control group
for the PMMA frame, a metal alloy frame was used, and
evaluated under the most unfavorable conditions. The
Mann-Whitney U-test followed by the Bonferroni correction was
used for statistical analysis. Results:
The span length significantly affected the fracture load.
Values ranged from 413 N for the 12 mm span length to 706 N
for the 7 mm span length. Thermal cycling and mechanical
loading significantly reduced fracture load values for FPDs
with the 12 mm span length, but there were no significant
effects for FPDs with 7 mm span length. Silicoating
pretreatment of the metal abutments significantly reduced
fracture load values. Replacement of the PMMA frame with a
metal frame increased fracture load values up to 1,075 N. (Am
J Dent 2007;20:405-410.)
Clinical significance:
Short span, inlay-retained composite FPDs with PMMA frames
showed acceptable fracture load values, but they cannot yet
be unreservedly recommended for clinical use.
*: Dr. Brigitte Ohlmann, Department of Prosthodontics MZK
II, University of Heidelberg, Im Neuenheimer Feld 400, 69120
Heidelberg, Germany.
E-*:
Brigitte_Ohlmann@med.uni-heidelberg.de
Effect of denture cleansers on the hardness of heat-
or auto-cured
acrylic- or silicone-based soft denture liner
s
Ayşe
Meşe,
dds
, phd
Abstract: Purpose:
To investigate the effect of short (24 hours) and long term (6
months) exposure to a variety of commercially available
denture cleanser solutions on the hardness of a variety of
acrylic- and silicone-based resilient liners that were either
heat- or auto-cured. Methods:
The denture liners investigated were an acrylic-based
heat-cured (Vertex Soft), acrylic-based auto-cured (Coe-Soft),
a silicon-based heat-cured (Molloplast-B) and silicon-based
auto-cured (Mollosil Plus) resilient liners. Cylindirical
specimens (20 mm diameter, 12 mm high) were made of each
material, using 10 replications for each test condition.
Immersion solutions consisted of distilled water (control),
and those based on alcohol, chlorhexidine, or an alkaline
peroxide (Dermacol, aqueous chlorhexidine, Steradent) as a
major active component. Specimens were fabricated according to
manufacturer directions. Shore-A durameter hardness readings
were taken on all specimens at each time interval and then
compared statistically using four-way ANOVA and Tukey HSD (α=
0.05). Results: The
results of this study indicated that, compared with distilled
water, significant effect on the hardness of the specimens
were found after immersion in all of the denture cleanser
solutions at 24 hours and at 6 months. Specimens immersed in
chlorhexidine showed higher and significant hardness changes
than those immersed in other cleanser solutions. Also, it was
determined that hardness of resilient liner materials
increased with time and significantly higher hardness was
recorded at 6 months for the auto-cured specimens compared
with their heat-cured counterparts. (Am
J Dent 2007;20:411-415).
Clinical significance:
The increasing hardness was greatest in acrylic resin-based,
auto-cured resilient liner in all cleansers especially in
chlorhexidine, which suggests that the use of this resilient
liner may not provide long-term clinical success.
*: Dr. Ayşe Meşe, Department of Prosthodontics,
Dental Faculty, The University of
Dicle Diyarbakır
,
Turkey
. E-*: amese@dicle.edu.tr
Resistance
to thermo-mechanical stress of different coupling agents
used
as intermediate layer in resin-fiber post bonds
Francesca Monticelli,
dds
,
msc
, phd,
Raquel Osorio,
dds
, phd, Franklin R.
Tay
, bdsc
(hons),
phd,
Fernanda T.
Sadek,
dds
,
msc
, phd,
Marco Ferrari,
md,
dds
, phd &
Manuel Toledano,
md,
dds
, phd
Abstract:
Purpose: To evaluate the microtensile bond strength (MTBS) of
different coupling agents used in fiber post-composite bonds
to withstand different in
vitro challenging procedures. Methods:
63 fiber posts (DT Light Post) etched with 10% hydrogen
peroxide were divided into three groups according to the
silane/adhesive system applied: (1) Porcelain Bond Activator
(PBA) + Clearfil SE Bond; (2) PBA + Clearfil Tri S Bond; (3)
Monobond-S. A composite build-up (Clearfil AP-X) was performed
around the post producing cylindrical specimens that were
divided into three subgroups according to the different aging
protocol: (1) 24-hour storage at room temperature; (2)
Thermocycling (5000 cycles, 5°/55°C dwell time: 30 seconds);
(3) Cyclic loading (45° angle, 20,000 cycles, load 5-50 N at
3.0 Hz). Samples were then cut obtaining sticks that were
loaded in tension until failure. Bond strength values were
statistically analyzed with two-way ANOVA and Tukey test (α
= 0.05). Failure mode was recorded and the morphologic aspect
of post/core interface after aging was evaluated under
SEM
. Results: Both
post superficial treatment, thermocycling and cyclic loading
influenced bond strength. After 24 hours, samples treated with
silane/adhesive couplings attained higher MTBS than those
bonded with conventional silane. No significant differences in
the microtensile bond strength at the post/core interface were
recorded between the different silane/adhesive couplings.
After challenging, no differences were found between the
tested groups. (Am J
Dent 2007;20: 416-420).
Clinical significance:
The application of a pre-activated single-bottle
MPS
silane solution formed relatively stable composite-fiber
post bonds, reducing time required for the clinical
procedure. Alternative coupling strategies may be desirable
allowing a more stable bonding mechanism.
*:
Dr. Francesca Monticelli, Department of Restorative Dentistry
and Dental Materials. Policlinico Le Scotte,
University
of
Siena
,
Viale Bracci
,
53100
Siena
,
Italy
.
E-*:
francescamonti@hotmail.it
|